Miernik stacjonarny RLC LCR-6002 10Hz ~ 2kHz / 0.05%

Zapytaj o produkt +
firma gwinstek

Opis

Features

  • 3.5” color LCD
  • 5 models (10Hz~2kHz/20kHz/100kHz/200kHz/300kHz)
  • Consecutive test frequency
  • Basic accuracy : 0.05%
  • Measuring speed up to 25ms (max.)
  • Full frequency range or spot OPEN/SHORT
  • 16 major/secondary parameter measurement combinations and two additional monitoring parameters(maximum four different parameters can be shown simultaneously)
  • DCR measurement and internal D.C. bias voltage (±2.5V)
  • PASS/FAIL judgment
  • Auto Level Control (ALC) function
  • BIN function provides 9BIN and 1AUX, totally 10 BIN
  • 10 steps listed tests to select different frequency, voltage and current criteria
  • Standard interface : RS-232C, Handler and USB storage
  • Compact size, ideal for automatic equipment (2U,1/2 RACK)
TEST FREQUENCY
LCR-6300: 10Hz?300kHz(±0.01%) (4 digits resolution)
LCR-6200: 10Hz?200kHz(±0.01%) (4 digits resolution)
LCR-6100: 10Hz?100kHz(±0.01%) (4 digits resolution)
LCR-6020: 10Hz? 20kHz(±0.01%) (4 digits resolution)
LCR-6002: 10Hz?  2kHz(±0.01%) (4 digits resolution)
OUTPUT IMPEDANCE
30Ω / 50Ω / 100Ω selectable
BASIC ACCURACY
Slow / Med 0.05%
Fast 0.1%
TEST SPEED
FAST:25ms / MED:100ms / SLOW:333ms
TEST SIGNAL LEVELS
AC Voltage 10.00mV- 2.00V (±10%)   CV :10.00mV- 2.00V(±6%)
     Current 100.0uA- 20.00mA (±10%) CC :100.0uA- 20.00mA(±6%) (@2VMax)
DCR ±1V(2Vpp), Square wave, 3Hz up 0.033A(Max)
DC BIAS
Internal ±2.5V (0.5%+0.005V)
DISPLAYRANGE
R, X, |Z|

G, B, |Y|

L

C

D

Q

θd

θr

DCR

Δ%

0.00001Ω ~ 99.9999MΩ

0.01nS ~ 999.999S

0.00001uH ~ 9999.99H

0.00001pF ~ 9999.99mF

0.00001 ~ 9.99999

0.00001 ~ 99999.9

-179.999° ~ 179.999°

-3.14159 ~ 3.14159

0.00001Ω ~ 99.9999MΩ

-99999% ~ 99999%

TEST MODE
16 Combinations Cs-Rs, Cs-D, Cp-Rp, Cp-D, Lp-Rp, Lp-Q, Ls-Rs, Ls-Q,
Rs-Q, Rp-Q, R-X, Z-θr, Z-θd, Z-D, Z-Q, DCR
Monitor parameter (2 selectable) Z, D, Q, Vac, Iac, Δ, Δ%, θr, θd, R, X, G, B, Y
LISTED MODE
10 steps
BIN FUNCTION
Comparator (9BIN,AUX:1BIN)
MEMORY
INT – panel setting 10 file name
INT – measured data 10000 Data(.csv)
USB storage 10 file name for setting, 9999 file name for data, 999 Log file for LCD screen
OTHER FUNCTION
Auto Level Control (ALC) ON/OFF
Average 1~256 times
Trigger INT / MAN / EXT / BUS
Delay 0ms~60s
Judgment PASS / FAIL
Auto LCZ ON / OFF
Screen Capture Saving in to USB (Bmp form)
DISPLAY
3.5’LCD, RGB color (320×240)
INTERFACE
RS-232(SCPI), Handler, USB Host
POWER SOURCE
AC 90V-250V, 50-60Hz, Max. 30VA
DIMENSION & WEIGHT
265(W)x107(H)x312(D) mm, Approx.3kg

 

Marka

GW Instek

firma gwinstek